6/9/2023 0 Comments Ets sv600 mindray![]() Then pull the expiration valve assembly out of the assembly horizontally. Rotate the expiration valve handwheel until the indicating arrow position.Water condensation in the expiration valve can be reduced by using a bacteria filter between the expiratory tube and expiration valveĮ: Unlocked state of the expiration valve If there is water build-up, empty it promptly. If there is water build-up inside the expiration valve, clear it before use.Ĭheck the expiratory water trap for water during the use of the ventilator. Whenthe patient’s exhaled gas arrives at the expiration valve, condensed water may appear at the expiration valve (including the expiratory flow sensor), compromising the measurement accuracy of expiratory flow sensor.Ĭheck the Sterilizable expiration valve assembly for water build-up when abnormal flow waveform or unstable tidal volume fluctuation is detected. ![]() The condensed water remains on the hose wall and finally enters the water trap. The patient’s exhaled warm and moist gas is condensed when it flows through the expiratory hose. Thus, the various components of the expiration module must be cleaned and disinfected before being reused. The Sterilizable expiration valve assembly is mainly responsible for functions such as pressure control, pressure monitoring and flow monitoring of the patient’s expiratory Philipilipase. Mindray SV300 SV600 SV800 E3 E5 ventilator sterilizable exhalation Expiration Valve Disinfected assembly EXP valve PN 115-021461-00